Bio MEMS

117602
Back to Bio MEMS

Module 2: Fundamentals of Microfluidic Systems

  1. Q1(a). Which of the following polymers is most commonly used in Soft Lithography for microfluidics? (i) PMMA (ii) PDMS (iii) Teflon (iv) Polycarbonate2025?m

    Module 2: Fundamentals of Microfluidic Systems

    Which of the following polymers is most commonly used in Soft Lithography for microfluidics?

    (i) PMMA
    (ii) PDMS
    (iii) Teflon
    (iv) Polycarbonate

    View this question on its own page →
    Worked Solution

    Answer: (ii) PDMS

    PDMS (Polydimethylsiloxane) is the most widely used polymer in soft lithography for microfluidics because it is transparent, flexible, biocompatible, gas-permeable, and easy to mold into microchannels.

  2. Q1(c). In microfluidic channels, the fluid flow is predominantly: (i) Turbulent (ii) Transitional (iii) Laminar (iv) Inviscid2025?m

    Module 2: Fundamentals of Microfluidic Systems

    In microfluidic channels, the fluid flow is predominantly:

    (i) Turbulent
    (ii) Transitional
    (iii) Laminar
    (iv) Inviscid

    View this question on its own page →
    Worked Solution

    Answer: (iii) Laminar

    Microfluidic flows generally have low Reynolds numbers because channel dimensions and flow velocities are small. Therefore, viscous forces dominate inertial forces and the flow is predominantly laminar, with little turbulent mixing.

  3. Q1(d). Which parameter is particularly important in determining the flow regime inside a microfluidic channel? (i) Reynolds number (ii) Atomic number (iii) Optical wavelength only (iv) Melting point20262m

    Module 2: Fundamentals of Microfluidic Systems

    Which parameter is particularly important in determining the flow regime inside a microfluidic channel?

    (i) Reynolds number
    (ii) Atomic number
    (iii) Optical wavelength only
    (iv) Melting point

    View this question on its own page →
    Worked Solution

    Answer: (i) Reynolds number

    Re = ρvD/μ compares inertial to viscous forces. In microchannels Re is usually far below 2000 (often below 1), so the flow is laminar.

  4. Q1(e). Which of the following is an important consideration when selecting a polymer for BioMEMS applications? (i) Biocompatibility (ii) Color alone (iii) Maximum bulk size (iv) Magnetic behavior only20262m

    Module 2: Fundamentals of Microfluidic Systems

    Which of the following is an important consideration when selecting a polymer for BioMEMS applications?

    (i) Biocompatibility
    (ii) Color alone
    (iii) Maximum bulk size
    (iv) Magnetic behavior only

    View this question on its own page →
    Worked Solution

    Answer: (i) Biocompatibility

    A BioMEMS polymer touches blood, cells or tissue, so it must not be toxic or trigger an immune response. Common examples are PDMS, PMMA and parylene.

  5. Q3(b). Discuss the motivation for developing Microfluidic Systems. Explain how the Surface-to-Volume ratio influences micro-scale phenomena.20257m

    Module 2: Fundamentals of Microfluidic Systems

    Discuss the motivation for developing Microfluidic Systems. Explain how the Surface-to-Volume ratio influences micro-scale phenomena.

    View this question on its own page →
    Worked Solution

    Motivation for Microfluidic Systems and Surface-to-Volume Ratio

    Microfluidics deals with controlling very small volumes of fluids, typically in channels with dimensions from micrometres to hundreds of micrometres.

    Motivation

    1. Very small sample and reagent consumption.
    2. Low operating cost.
    3. Rapid analysis and reaction times.
    4. High-throughput parallel processing.
    5. Integration of sample preparation, reaction and detection.
    6. Portable point-of-care diagnostic systems.
    7. Precise control of fluid transport and cellular environments.

    Surface-to-volume ratio

    For a characteristic length LL, surface area scales approximately as L2L^2 while volume scales as L3L^3. Therefore, surface-to-volume ratio scales as 1/L1/L and increases strongly as dimensions decrease.

    Consequences include:

    • surface forces and adhesion become important,
    • heat transfer becomes rapid,
    • diffusion can dominate mass transport,
    • capillary forces become significant,
    • wall interactions strongly influence fluid behavior.

    Thus, the high surface-to-volume ratio is one of the main reasons microfluidic systems behave differently from conventional fluidic systems.

  6. Q4(a). Describe the different types of microfluidic channels and their working principles.20257m

    Module 2: Fundamentals of Microfluidic Systems

    Describe the different types of microfluidic channels and their working principles.

    View this question on its own page →
    Worked Solution

    Types of Microfluidic Channels and Working Principles

    Microfluidic channels can be classified according to geometry and function.

    1. Straight channels

    A simple rectangular or circular passage transports fluid from inlet to outlet. At low Reynolds number, flow is mainly laminar and pressure-driven.

    2. Serpentine channels

    The channel contains repeated bends. The longer path increases residence time and can enhance mixing through repeated deformation of fluid streams.

    3. T-junction channels

    Two streams meet at a T-shaped junction. They are commonly used for controlled mixing and droplet generation.

    4. Y-junction channels

    Two inlet streams merge at an angle into one channel. They are useful for controlled co-flow and diffusion-based mixing.

    5. Flow-focusing channels

    A central stream is squeezed by side streams, allowing controlled formation of droplets or particles.

    6. Microchambers

    Expanded regions connected to channels provide space for reactions, cell trapping, sensing or incubation.

    Working principle

    Fluid movement is produced by pressure-driven flow, electroosmotic flow or other microscale actuation methods. Because flow is generally laminar, mixing is often achieved by diffusion or deliberate channel geometry rather than turbulence.

  7. Q4(a). Explain the working principle of a microfluidic channel with a suitable schematic diagram.20267m

    Module 2: Fundamentals of Microfluidic Systems

    Explain the working principle of a microfluidic channel with a suitable schematic diagram.

    View this question on its own page →
    Worked Solution

    Working principle of a microfluidic channel

    A microfluidic channel is a tiny conduit, typically 10 to 500 µm wide, etched or moulded in silicon, glass or a polymer and sealed with a cover. It carries nanolitre to microlitre volumes of liquid from an inlet to an outlet.

    SVGVector Architecture Diagram (SVG)
    Microfluidic channel (side view, laminar flow)Cover (glass / PDMS)Substrate (silicon / glass)InletOutletParabolic velocity profile: zero at walls, maximum at centre

    Working principle

    1. Driving force: a syringe pump, pressure source, electric field (electroosmosis) or capillary action pushes the liquid through the channel.
    2. Laminar flow: the Reynolds number Re = ρvD_h/μ is small (usually well below 100 and often below 1), so streams move in parallel layers without turbulence. Mixing happens only by diffusion.
    3. Velocity profile: pressure-driven flow is parabolic: zero at the walls (no-slip) and maximum at the centre.
    4. Hydraulic resistance: flow rate Q = ΔP/R_h. For a circular channel R_h = 8μL/(πr⁴); for a rectangular channel it rises sharply as the height shrinks, so very small channels need high pressure.
    5. Surface effects: surface tension, capillarity and wall charge dominate at this scale.

    Key relations

    Quantity Expression
    Reynolds number Re = ρvD_h/μ
    Hydraulic diameter D_h = 4A/P (A = area, P = wetted perimeter)
    Flow rate Q = ΔP/R_h
    Diffusive mixing time t ≈ w²/D

    Exam tip

    Draw the labelled side view, show the parabolic profile, and state that the flow is laminar because Re is small.

  8. Q4(b). What are biocompatible polymers? Discuss the properties and applications of PDMS (Polydimethylsiloxane) in BioMEMS.20257m

    Module 2: Fundamentals of Microfluidic Systems

    What are biocompatible polymers? Discuss the properties and applications of PDMS (Polydimethylsiloxane) in BioMEMS.

    View this question on its own page →
    Worked Solution

    Biocompatible Polymers and PDMS in BioMEMS

    Biocompatible polymers are polymeric materials that can perform their intended function in contact with biological tissues, cells or fluids without causing unacceptable adverse biological responses.

    Important properties of PDMS

    • Biocompatibility for many laboratory and diagnostic applications.
    • Optical transparency, useful for microscopy.
    • Flexible and elastomeric.
    • Easy replication by soft lithography.
    • Low cost and rapid prototyping.
    • Good gas permeability, especially oxygen and carbon dioxide.
    • Low surface energy and easy molding.
    • Can be bonded to glass or other substrates after surface treatment.

    Applications

    1. Microfluidic channels.
    2. Cell culture and cell manipulation platforms.
    3. Lab-on-Chip devices.
    4. Drug delivery research.
    5. Micropumps and valves.
    6. Diagnostic and biological analysis devices.

    Limitation

    PDMS can absorb some small hydrophobic molecules and its surface properties can change with treatment, so material selection must match the biomedical application.

  9. Q4(b). Discuss different types of microfluidic channels and explain their operating principles.20267m

    Module 2: Fundamentals of Microfluidic Systems

    Discuss different types of microfluidic channels and explain their operating principles.

    View this question on its own page →
    Worked Solution

    Types of microfluidic channels and their operating principles

    Microchannels are classified by geometry and by how the fluid is driven.

    By geometry and function

    Type Description Use
    Straight channel Simple uniform path Reaction, detection, separation
    Serpentine (meander) Long folded path in a small area Mixing, long residence time, PCR
    T- and Y-junction Two inlets merge into one outlet Laminar mixing, droplet generation
    Flow-focusing Sheath flows squeeze the core stream Droplet generation, cell focusing
    Spiral Curved path creates Dean flow Inertial cell separation
    Branched / tree Splits flow into parallel channels Distribution, gradient generators

    By driving principle

    Type Operating principle
    Pressure-driven A pump or syringe applies ΔP; parabolic (Poiseuille) profile, Q = ΔP/R_h
    Electrokinetic (electroosmotic) An electric field moves the ionic layer along charged walls; nearly flat (plug) profile
    Capillary-driven Surface tension pulls liquid into hydrophilic channels; no external pump
    Centrifugal Rotation of a disc pushes fluid outward; valves set by spin speed

    By cross-section

    Rectangular (common from soft lithography), trapezoidal or V-shaped (anisotropic KOH etching of silicon), semicircular (isotropic etching of glass), circular (capillary tubes).

    MermaidDiagram
    Rendering diagram…

    Exam tip

    Two tables (geometry and driving force) plus the hydraulic-resistance idea give full marks.