Explain the electrode–skin interface and its significance in biosignal acquisition.

Worked SolutionAI Assisted

An electrode converts ionic current in tissue into electronic current in the measuring circuit. At the electrode–electrolyte interface, charge transfer and double-layer effects create a half-cell potential, resistance, and capacitance; skin adds further resistance/capacitance. Dry or loose electrodes have higher impedance and motion can change it, producing artefacts. High or mismatched impedance increases noise susceptibility and can convert common-mode interference into differential noise. Improve performance by cleaning/preparing skin, using suitable Ag–AgCl electrodes and gel, securing leads, and using a high-input-impedance amplifier with high CMRR. Medical electrical isolation and current-limiting protection are essential.