2025 question paper
Biomedical Transducer and Instrumentation
25 questions
Q1(a). The process of converting physiological energy into an electrical signal is called: (i) Amplification (ii) Transduction (iii) Modulation (iv) Filtering20252m
Module 1: Biomedical transducers and sensors
View this question on its own page →The process of converting physiological energy into an electrical signal is called:
(i) Amplification
(ii) Transduction
(iii) Modulation
(iv) FilteringWorked SolutionCorrect answer: (ii) Transduction. Transduction converts a physiological quantity (pressure, temperature, displacement, chemical concentration, or bioelectric activity) into a measurable signal, commonly electrical. Amplification increases magnitude; modulation varies a carrier parameter; filtering selects frequencies.
Q1(b). Which sensor is commonly used to measure blood oxygen saturation? (i) Thermistor (ii) Pulse oximeter (iii) Strain gauge (iv) Piezoelectric sensor20252m
Module 2: Biomedical sensors and measurements
View this question on its own page →Which sensor is commonly used to measure blood oxygen saturation?
(i) Thermistor
(ii) Pulse oximeter
(iii) Strain gauge
(iv) Piezoelectric sensorWorked SolutionCorrect answer: (ii) Pulse oximeter. It estimates arterial oxygen saturation using red and infrared light, absorbed differently by oxyhaemoglobin and deoxyhaemoglobin. The pulsatile optical component helps estimate SpO₂ non-invasively.
Q1(c). ECG signals primarily represent the electrical activity of: (i) Brain (ii) Lungs (iii) Skeletal muscles (iv) Heart20252m
Module 3: Biomedical signal acquisition
View this question on its own page →ECG signals primarily represent the electrical activity of:
(i) Brain
(ii) Lungs
(iii) Skeletal muscles
(iv) HeartWorked SolutionCorrect answer: (iv) Heart. ECG records body-surface voltage differences produced by cardiac depolarisation and repolarisation. The P wave reflects atrial depolarisation, QRS ventricular depolarisation, and T ventricular repolarisation.
Q1(d). The main function of signal conditioning circuits is to: (i) Generate signals (ii) Display signals (iii) Improve signal quality (iv) Store signals20252m
Module 3: Biomedical signal acquisition
View this question on its own page →The main function of signal conditioning circuits is to:
(i) Generate signals
(ii) Display signals
(iii) Improve signal quality
(iv) Store signalsWorked SolutionCorrect answer: (iii) Improve signal quality. Signal conditioning prepares sensor output through amplification, filtering, isolation, impedance matching, linearisation, and level shifting, improving signal-to-noise ratio before display or digitisation.
Q1(e). Which electrode is most commonly used for surface ECG recording? (i) Ag–AgCl electrode (ii) Glass electrode (iii) Needle electrode (iv) Microelectrode20252m
Module 3: Biomedical signal acquisition
View this question on its own page →Which electrode is most commonly used for surface ECG recording?
(i) Ag–AgCl electrode
(ii) Glass electrode
(iii) Needle electrode
(iv) MicroelectrodeWorked SolutionCorrect answer: (i) Ag–AgCl electrode. Silver/silver-chloride electrodes provide stable, low-polarisation potentials and low noise for surface ECG when paired with proper skin preparation and conductive gel.
Q1(f). The common mode rejection ratio (CMRR) is a key parameter of: (i) Power supply (ii) Instrumentation amplifier (iii) Oscillator (iv) ADC20252m
Module 3: Biomedical signal acquisition
View this question on its own page →The common mode rejection ratio (CMRR) is a key parameter of:
(i) Power supply
(ii) Instrumentation amplifier
(iii) Oscillator
(iv) ADCWorked SolutionCorrect answer: (ii) Instrumentation amplifier. CMRR measures an amplifier’s ability to reject common-mode input: CMRR = Ad/Acm, and CMRR(dB) = 20 log₁₀(Ad/Acm). High CMRR is valuable for rejecting common mains interference on both ECG leads.
Q1(g). MRI imaging is based on the principle of: (i) X-ray absorption (ii) Ultrasonic reflection (iii) Radioactive decay (iv) Nuclear magnetic resonance20252m
Module 5: Medical imaging systems
View this question on its own page →MRI imaging is based on the principle of:
(i) X-ray absorption
(ii) Ultrasonic reflection
(iii) Radioactive decay
(iv) Nuclear magnetic resonanceWorked SolutionCorrect answer: (iv) Nuclear magnetic resonance. MRI excites nuclei, mainly hydrogen, in a strong magnetic field using radio-frequency pulses. Relaxation signals are spatially encoded with gradients and reconstructed into images; MRI does not use ionising X-rays.
Q1(h). Which imaging modality uses ionizing radiation? (i) Ultrasound (ii) CT scan (iii) MRI (iv) Thermography20252m
Module 5: Medical imaging systems
View this question on its own page →Which imaging modality uses ionizing radiation?
(i) Ultrasound
(ii) CT scan
(iii) MRI
(iv) ThermographyWorked SolutionCorrect answer: (ii) CT scan. CT uses ionising X-rays and detectors to measure attenuation from multiple angles, then reconstructs cross-sectional images. Ultrasound uses sound waves; MRI uses magnetic fields and RF energy.
Q1(i). A pacemaker is primarily used to treat: (i) Hypertension (ii) Respiratory failure (iii) Cardiac arrhythmias (iv) Epilepsy20252m
Module 6: Therapeutic and assistive biomedical devices
View this question on its own page →A pacemaker is primarily used to treat:
(i) Hypertension
(ii) Respiratory failure
(iii) Cardiac arrhythmias
(iv) EpilepsyWorked SolutionCorrect answer: (iii) Cardiac arrhythmias. A pacemaker delivers controlled electrical impulses when the heart’s intrinsic rate or conduction is inadequate, helping maintain an appropriate rhythm.
Q1(j). Wearable biosensors are mainly associated with: (i) Invasive surgery (ii) Medical imaging (iii) Radiation therapy (iv) Remote health monitoring20252m
Module 6: Therapeutic and assistive biomedical devices
View this question on its own page →Wearable biosensors are mainly associated with:
(i) Invasive surgery
(ii) Medical imaging
(iii) Radiation therapy
(iv) Remote health monitoringWorked SolutionCorrect answer: (iv) Remote health monitoring. Wearable biosensors track variables such as heart rate, activity, temperature, and oxygen saturation and transmit trends to a phone or clinical platform. Data quality and privacy remain important.
- Install PocketlyBudget App•POCKETLY
Pocketly on Android
Log daily food, travel, and hostel expenses instantly. Set monthly spending limits and see where your money goes without complex spreadsheets.
Q2(a). Explain the basic principles of biomedical transducers.20257m
Module 1: Biomedical transducers and sensors
View this question on its own page →Explain the basic principles of biomedical transducers.
Worked SolutionPrinciple: A biomedical transducer converts a physiological measurand—such as pressure, temperature, flow, displacement, chemical concentration, or bioelectric potential—into a usable output, usually electrical. The measurand acts on a sensing element, changing resistance, capacitance, inductance, charge, voltage, or optical intensity. The output is conditioned (amplified/filtered), digitised if needed, then displayed or stored. Examples: thermistor (temperature/resistance), strain gauge (force/pressure/resistance), piezoelectric sensor (stress/charge), ECG electrode (ionic-to-electronic interface), and photodiode in pulse oximetry. Key characteristics include sensitivity, range, linearity, accuracy, resolution, repeatability, response time, drift, hysteresis, and noise. Medical designs also require electrical safety and, where applicable, biocompatibility.
Q2(b). Classify biomedical sensors with suitable examples.20257m
Module 1: Biomedical transducers and sensors
View this question on its own page →Classify biomedical sensors with suitable examples.
Worked SolutionClassification of biomedical sensors: (1) By measurand: temperature (thermistor), pressure (strain-gauge sensor), flow (ultrasonic flowmeter), displacement (LVDT), force (load cell), bioelectric potential (ECG electrode), chemical variable (pH sensor). (2) By transduction: resistive, capacitive, inductive, piezoelectric, thermoelectric, optical, electrochemical, semiconductor. (3) By energy: active/self-generating (piezoelectric, thermocouple) versus passive/excitation-required (strain gauge, RTD). (4) By contact: contact (skin electrode) versus non-contact (infrared thermometer). (5) By invasiveness: non-invasive (pulse oximeter) versus invasive (catheter pressure sensor). Selection depends on range, accuracy, bandwidth, sensitivity, response time, patient safety, artefacts, cost, and environment.
Q3(a). Describe the physiological origin of ECG signals with a neat diagram. List the applications of ECG in clinical diagnosis.20257m
Module 2: Biomedical signals
View this question on its own page →Describe the physiological origin of ECG signals with a neat diagram. List the applications of ECG in clinical diagnosis.
Worked SolutionOrigin: Cardiac electrical activity begins at the SA node, spreads through the atria, pauses at the AV node, then travels through the His–Purkinje system to activate the ventricles. Depolarisation and repolarisation create fields conducted through body tissues; surface electrodes measure their potential differences. ECG waveform: P = atrial depolarisation; QRS = ventricular depolarisation; T = ventricular repolarisation (atrial repolarisation is usually hidden by QRS). Recording chain: electrodes → differential/instrumentation amplifier → protection/isolation → filters → ADC/display. Applications: rhythm and conduction assessment, arrhythmia detection, evidence of ischaemia/infarction, monitoring during surgery/critical care, assessing drug/electrolyte effects, and tracking treatment. ECG must be interpreted with clinical context; a normal trace does not exclude all cardiac disease.
Q3(b). An ECG amplifier has a differential gain of 1000 and a common-mode gain of 0.1. Calculate the CMRR in dB.20257m
Module 3: Biomedical signal acquisition
View this question on its own page →An ECG amplifier has a differential gain of 1000 and a common-mode gain of 0.1. Calculate the CMRR in dB.
Worked SolutionGiven Ad = 1000 and Acm = 0.1. Linear CMRR = Ad/Acm = 1000/0.1 = 10,000. In decibels, CMRR(dB) = 20 log₁₀(10,000) = 20 × 4 = 80 dB. Thus CMRR is 10,000 (linear), or 80 dB; higher CMRR means stronger rejection of common-mode noise.
Q4(a). Explain the electrode–skin interface and its significance in biosignal acquisition.20257m
Module 3: Biomedical signal acquisition
View this question on its own page →Explain the electrode–skin interface and its significance in biosignal acquisition.
Worked SolutionAn electrode converts ionic current in tissue into electronic current in the measuring circuit. At the electrode–electrolyte interface, charge transfer and double-layer effects create a half-cell potential, resistance, and capacitance; skin adds further resistance/capacitance. Dry or loose electrodes have higher impedance and motion can change it, producing artefacts. High or mismatched impedance increases noise susceptibility and can convert common-mode interference into differential noise. Improve performance by cleaning/preparing skin, using suitable Ag–AgCl electrodes and gel, securing leads, and using a high-input-impedance amplifier with high CMRR. Medical electrical isolation and current-limiting protection are essential.
Q4(b). Compare surface electrodes and needle electrodes.20257m
Module 3: Biomedical signal acquisition
View this question on its own page →Compare surface electrodes and needle electrodes.
Worked SolutionSurface vs needle electrodes: Surface electrodes sit on skin, are non-invasive and comfortable, and record activity summed over a wider region; used for ECG, routine EEG, and surface EMG. Needle electrodes enter tissue, are invasive and may cause discomfort, but record more localised intramuscular activity; used in diagnostic needle EMG. Surface recording has lower spatial selectivity and is affected by skin impedance/motion. Needle recording requires sterile technique and carries risks such as bleeding, infection, and tissue injury.
Q5(a). Discuss amplification and filtering techniques used in biomedical signal processing.20257m
Module 3: Biomedical signal acquisition
View this question on its own page →Discuss amplification and filtering techniques used in biomedical signal processing.
Worked SolutionAmplification: Biomedical signals are often very small, so low-noise instrumentation amplifiers provide differential gain, high input impedance, and high CMRR. Gain must avoid saturation from offsets or motion artefacts; isolation and patient protection are required. Filtering: high-pass filters suppress baseline wander; low-pass filters attenuate high-frequency noise; notch filters reduce 50/60 Hz mains interference; band-pass filters retain a chosen physiological band. Filters can be analogue before ADC and digital afterward. Use an anti-aliasing low-pass filter before sampling and sample above twice the highest retained frequency. Excessive filtering can distort diagnostically important waveform morphology.
Q5(b). A biomedical signal has a frequency range of 0.05 Hz to 150 Hz. Design suitable high-pass and low-pass filter cutoff frequencies for proper signal acquisition.20257m
Module 3: Biomedical signal acquisition
View this question on its own page →A biomedical signal has a frequency range of 0.05 Hz to 150 Hz. Design suitable high-pass and low-pass filter cutoff frequencies for proper signal acquisition.
Worked SolutionThe desired band is 0.05–150 Hz. A nominal design uses a high-pass cutoff near 0.05 Hz and a low-pass cutoff near 150 Hz (choose a little wider if flat passband at both edges is required). For first-order RC filters, fc = 1/(2πRC). Example high-pass: C = 10 μF gives R = 1/(2π×0.05×10 μF) ≈ 318 kΩ. Example low-pass: C = 0.1 μF gives R = 1/(2π×150×0.1 μF) ≈ 10.6 kΩ. Add anti-alias filtering and sample above 300 samples/s, preferably with practical margin. Actual order and cutoff depend on ripple, attenuation, and signal type.
Q6(a). Explain the working principle and instrumentation of X-ray imaging.20257m
Module 5: Medical imaging systems
View this question on its own page →Explain the working principle and instrumentation of X-ray imaging.
Worked SolutionPrinciple: An X-ray tube emits photons through the patient; tissues attenuate the beam differently according to thickness, density, and composition. A detector measures transmitted intensity to form a projection image. Instrumentation: high-voltage supply, cathode/anode X-ray tube, collimator, patient table, anti-scatter grid where used, digital detector, console/computer, shielding, and safety interlocks. Heated cathode electrons accelerate toward the anode and generate X-rays; the collimated beam passes through the body and the detector converts it to image data. Bone usually appears more radiopaque than soft tissue; air attenuates least. Advantages: fast, accessible, excellent for bones/chest. Limitations: ionising radiation, overlapping structures, and limited soft-tissue contrast. Justification, dose optimisation, collimation, and shielding are important.
Q6(b). Compare CT scan and MRI.20257m
Module 5: Medical imaging systems
View this question on its own page →Compare CT scan and MRI.
Worked SolutionCT vs MRI: CT uses X-ray attenuation measured from many angles and computer reconstruction; MRI uses nuclear magnetic resonance in a strong magnetic field with RF pulses and gradients. CT is usually faster and excellent for trauma, lung, bone, and acute bleeding, but uses ionising radiation. MRI avoids ionising radiation and often gives superior soft-tissue detail (brain, spinal cord, joints), but is slower, costlier, motion-sensitive, and requires screening for incompatible implants/metal. CT contrast is often iodinated; MRI contrast may be gadolinium-based when indicated. Choice depends on clinical need, urgency, contraindications, and risk-benefit.
- Install PocketlyBudget App•POCKETLY
Pocketly on Android
Log daily food, travel, and hostel expenses instantly. Set monthly spending limits and see where your money goes without complex spreadsheets.
Q7(a). Describe the construction and working of a cardiac pacemaker.20257m
Module 6: Therapeutic and assistive biomedical devices
View this question on its own page →Describe the construction and working of a cardiac pacemaker.
Worked SolutionConstruction: pulse generator with battery, timing/control electronics and output stage; insulated leads and electrodes (or a leadless intracardiac unit). Working: a demand pacemaker senses intrinsic cardiac activity. If an adequate beat occurs within the programmed interval, it inhibits a pulse; if not, it delivers a controlled electrical stimulus through the lead to depolarise myocardium and trigger contraction. Rate-responsive devices adapt pacing to sensed activity. Block: battery → controller/timer → output stage → lead/electrode → myocardium, with sensed cardiac signal fed back to the controller. Uses include symptomatic bradycardia and conduction disorders. Follow-up checks battery, leads, sensing, capture thresholds, and stored rhythm. Risks include infection and lead displacement; specialist implantation/programming is required.
Q7(b). An ultrasound system operates at a frequency of 5 MHz. If the velocity of sound in human tissue is 1540 m/s, calculate the wavelength of the ultrasound wave.20257m
Module 5: Medical imaging systems
View this question on its own page →An ultrasound system operates at a frequency of 5 MHz. If the velocity of sound in human tissue is 1540 m/s, calculate the wavelength of the ultrasound wave.
Worked SolutionGiven f = 5 MHz = 5×10⁶ Hz and sound speed v = 1540 m/s. Wavelength λ = v/f = 1540/(5×10⁶) = 3.08×10⁻⁴ m = 0.308 mm. Higher frequency produces shorter wavelength and usually better spatial resolution, but attenuation through tissue also increases.
Q8(a). Discuss emerging trends in wearable health monitoring systems.20257m
Module 6: Therapeutic and assistive biomedical devices
View this question on its own page →Discuss emerging trends in wearable health monitoring systems.
Worked SolutionWearable monitoring trends include continuous sensing with watches, rings, patches and smart textiles; miniaturised low-power electronics; remote monitoring via phones/cloud; multimodal sensors (ECG, photoplethysmography, motion and temperature); AI-assisted trend detection; and personalised long-term monitoring for chronic care and rehabilitation. Benefits include early alerts and longitudinal data. Challenges include accuracy across skin tones/activity, motion artefacts, calibration, false alarms, battery life, interoperability, cybersecurity, consent, privacy, and clinical validation. Wearables support but do not automatically replace clinical-grade assessment.
Q8(b). Explain regulatory standards and ethical considerations in biomedical instrumentation.20257m
Module 7: Biomedical standards and ethics
View this question on its own page →Explain regulatory standards and ethical considerations in biomedical instrumentation.
Worked SolutionBiomedical devices are governed by requirements for safety, essential performance, quality, and risk management. Examples include ISO 13485 (quality management), ISO 14971 (risk management), and the IEC 60601 series (medical electrical equipment); applicability depends on device and jurisdiction. Engineering practice includes hazard analysis, verification/validation, electrical leakage and EMC testing, biocompatibility where relevant, traceability, cybersecurity, and post-market surveillance. Ethics includes informed consent, privacy, secure health-data handling, minimising harm, fairness across populations, transparency about limitations, accessibility, and responsible AI use. Check current local regulations and the standards applicable to the intended device.
Q9. Write short notes on any two of the following (7 marks each): (a) EMG (b) Artifact removal techniques (c) Biomedical robotics and artificial intelligence (d) Nanotechnology in biomedical instrumentation.202514m
Module 7: Emerging technologies in biomedical instrumentation
View this question on its own page →Write short notes on any two of the following (7 marks each):
(a) EMG
(b) Artifact removal techniques
(c) Biomedical robotics and artificial intelligence
(d) Nanotechnology in biomedical instrumentation.Worked Solution(a) EMG: Records skeletal-muscle electrical activity. Surface EMG is non-invasive; needle EMG provides local intramuscular recordings. Applications include neuromuscular diagnosis, rehabilitation, prosthetic control, and human–machine interfaces. Limitations include motion artefacts, electrode-placement sensitivity, and cross-talk. (b) Artefact removal: Common artefacts include mains hum, baseline drift, motion, electrode pops, and muscle noise. Remedies include skin preparation, shielding, differential amplification, notch/band-pass filtering, adaptive filtering, wavelets, and quality checks; avoid removing true physiology. (c) Biomedical robotics and AI: Used in surgery, rehabilitation, assistive devices, laboratory automation, image segmentation, and motion planning. Benefits are precision and repeatability; risks include safety, bias, explainability, cybersecurity, and need for clinical validation/oversight. (d) Nanotechnology: Nanomaterials can improve biosensor sensitivity, imaging contrast, drug delivery, and implant coatings. Challenges include toxicity, biocompatibility, long-term fate, reproducibility, manufacturing quality, and regulation. The exam asks any two, 7 marks each.