Module 3: Biomedical signal acquisition

20257m

A biomedical signal has a frequency range of 0.05 Hz to 150 Hz. Design suitable high-pass and low-pass filter cutoff frequencies for proper signal acquisition.

Worked SolutionAI Assisted

The desired band is 0.05–150 Hz. A nominal design uses a high-pass cutoff near 0.05 Hz and a low-pass cutoff near 150 Hz (choose a little wider if flat passband at both edges is required). For first-order RC filters, fc = 1/(2πRC). Example high-pass: C = 10 μF gives R = 1/(2π×0.05×10 μF) ≈ 318 kΩ. Example low-pass: C = 0.1 μF gives R = 1/(2π×150×0.1 μF) ≈ 10.6 kΩ. Add anti-alias filtering and sample above 300 samples/s, preferably with practical margin. Actual order and cutoff depend on ripple, attenuation, and signal type.